Intins - Ocean Optics' Exclusive Official Distributor in Vietnam
  • 02432045963
  • sales@intins.vn

DH-3P-BAL-CAL

Khác HF Corp

Overview

 

Radiometrically calibrated light sources from Ocean Insight are used to calibrate the absolute spectral response of a spectrometer system. Using these light sources and our software, you can determine absolute intensity values across UV, Visible and NIR wavelengths. 

Radiometrically calibrated light sources are carefully characterized to deliver a known quantity of light with very low uncertainty. Each source is measured to NIST-traceable standards, creating an unbroken chain of traceability and providing you with the highest quality and most reliable data.

 

 

  • Stable sources -- provide consistent output for the most reliable data
  • Multiple options -- radiometrically calibrated sources for UV-Visible-NIR and Visible-NIR wavelengths
  • Digital calibration data -- can be loaded into software for absolute intensity data
  • Applications flexibility -- measure LEDs, digital displays and other radiant sources
 
 

 

Engineering

Nominal Bulb Power

25 W (deuterium), 20 W (tungsten halogen)

Wavelength Range

230nm - 1.1μm

Calibrated For

Absolute irradiance (µW/cm2/nm)

Calibration Lifetime

50 hours

Drift (Voltage)

<0.01% per hour

Minimum Uncertainty Level

4.0% at 700nm

Stability (Current)

<5 x 10-6 peak-to-peak (0.1-10.0 Hz)

Warm-up Time

40 minutes (deuterium),

20 minutes (tungsten halogen)

Environmental

Safety & Regulatory

CE, ROHS, WEEE

Operating Humidity Specification:

5-95% without condensation at 40°C

Operating Temperature

5 °C – 35 °C

Physical

Fiber Connector

SMA 905; CC-3-UV-S or 6.35 mm barrel for cosine corrector

Shutter

Yes

Dimensions

15 x 13.5 x 28.5 cm

Power Consumption

Approximately 78VA

Power Requirements

85-264 V, 50/60 Hz

Weight

6 kg